How To Use Amprobe Am-530 For Current Draw Test On A Car
BEHA-AMPROBE AM-530-EUR Digital Multimeter User Manual
AM-520-EUR / AM-530-EUR Digital Multimeter
- Wink low-cal
- LCD Brandish
- Part Buttons
- Flash light Button
- Rotary Switch
- SELECT Button
- Input Terminal for voltage, diode, capacitance, Resistance, continuity and temperature measurement
- COM (render) final for all measurements
- Input Terminal for Air-conditioning/DC mA/uA measurement
- Input Last for Air-conditioning/DC A measurement to10A
Screen Display
- The Meter selects the range with all-time resolution
- Direct Current
- Negative reading
- Alternating Current
- Low battery indicator
- Information hold
- Diode exam
- Continuity test
- Relative zero mode
- Non-Contact Voltage
- Measurement units for Resistance
- Measurement units for Frequency
- Measurement units for Voltage
- Measurement units for Electric current
- Measurement units for Capacitance
- Machine Power Off
- Maximum / minimum reading memory
- Measurement unit of measurement for Temperature
- Analog bar graph display
SYMBOLS
| Caution ! Take a chance of electric stupor. |
| Caution! Refer to the explanation in this Manual |
| | Alternate Current (AC) |
| | Direct Current (DC) |
| The equipment is protected by double insulation or reinforced insulation |
| Earth (Footing) |
| Audible tone |
| Bombardment |
| Complies with European Directives |
| Conforms to relevant Australian standards |
| Canadian Standards Association (NRTL/C) |
| Practice not dispose of this product as unsorted municipal waste. Contact a qualified recycler. |
Condom Information
The Meter complies with:
IEC/EN 61010-1 3rd Edition, UL61010-ane 2nd Ed. and Can/CSA C22.2 No.
61010.1-0.92 to Category 3 600 Volts, Pollution degree 2
IEC/EN 61010-two-030
IEC/EN 61010-ii-31 for examination leads
EMC IEC/EN 61326-i
Measurement Category Iii (True cat III) is for measurements performed in the edifice installation. Examples are measurements on distribution boards, circuit- breakers, wiring, including cables, motorbus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for instance, stationary motors with permanent connectedness to the fixed installation.
Measurement Category2 (True cat Ii) iis for measurements performed on circuit directly continued to low voltage installation. Examples are measurements on household appliances, portable tools and similar equipments.
Warning: Read Before Using
- To avoid possible electrical daze or personal injury, follow these instructions and use the Meter only equally specified in this manual.
- Do not use the Meter or examination leads if they appear damaged, or if the Meter is not operating properly. If in dubiousness, take the Meter serviced.
- Always use the proper part and range for measurements.
- Earlier rotating the function range selection switch, disconnect test probe from excursion under test.
- Verify the Meter's operation by measuring on a known voltage source.
- Do not apply more than the rated voltage, as marked on the Meter, between the test probe or between any test probe and earth ground.
- Utilize the Meter with caution for voltages above 30 Vac rms, 42 Vac tiptop, or 60 Vdc. These voltages pose electric stupor hazards.
- Disconnect circuit power and discharge all high-voltage capacitors before testing resistance.
- Practice not use the Meter effectually explosive gas or vapor.
- When using the test leads, keep your fingers behind the finger guards.
- Remove examination leads from the Meter before opening the Meter case or battery door.
UNPACKING AND INSPECTION
Your shipping carton should include:
- 1 AM-520-EUR or AM-530-EUR
- 1 Pair of examination leads
- 1 Temperature probe
- 1 Velcro strap
- i 9V (6F22) bombardment (installed)
- 1 User manual
- 1 Carrying case
If any of the items are damaged or missing, return the complete package to the place of purchase for an exchange.
FEATURES
AM-520-EUR is designed for HVAC applications with cardinal functions such as temperature, micro amps used for flame sensor troubleshooting, as well as and capacitance to check the motor startup capacitors. The AM-520-EUR measures a complete range of electrical parameters and features a built in flashlight, a "third paw" probe holder and VoltTect non-contact voltage detection. Safety rated to True cat III 600V.
AM-530-EUR is the fully-featured multimeter of pick for the professional electrical contractor. Measure out and verify presence of voltage in order to connect equipment or to perform repairs, run new wiring, bank check continuity of electrical connections, identify diddled fuses, troubleshoot motors or check transformers. The AM-530-EUR features Truerms sensing to accurately measure voltage on systems affected by harmonics, a congenital in flashlight to detect wire colors in the dark, a "third mitt" probe holder and non-contact voltage detection. Safety rated to Cat 3 600V.
- Measurements: Voltage up to 600VAC and 600VDC, Ac/DC current, Resistance, Frequency, Capacitance, Temperature.
- Frequency, Capacitance, Duty Cycle for troubleshooting applications
- Special Functions:
- Not-contact Voltage Detection
- Audible continuity
- Diode Test
- Backlit LCD brandish with analog bar graph
- Events:
- Data hold
- MAX / MIN Retention
- Relative zero manner
- Congenital in work light (flashlight)
- Built in test leads storage and "3rd hand holder"
- Auto and Transmission ranging
- Auto ability off
- Low battery alarm
- Velcro strap to hang a meter
- Safety: CAT Iii 600V
MAKING MEASUREMENT
-
Use the proper function and range for measurements. - To avert possible electric daze, personal injury or amercement to the
Meter, disconnect circuit ability and discharge all high-voltage capacitors
before testing resistance and diode. - Connecting examination leads:
- Connect the common (COM) test lead to the excursion before connecting the live lead;
• Afterwards measurement, remove alive atomic number 82 before removing the common (COM) examination lead from the circuit
- Connect the common (COM) test lead to the excursion before connecting the live lead;
- Symbol "OL" is displayed on LCD when the measurement is out of range.
Rotary Switch Positions
| Switch Position | Measurement Function |
| Air-conditioning or DC voltage measurement (employ SELECT button for switching to Air-conditioning or DC). |
| Ω | Resistance measurement |
| | Voltage measurement of diode PN junction |
| | Continuity measurement |
| | Capacitance measurement |
| HZ | Frequency measurement |
| °C °F | Temperature measurement |
| NCV | Non-contact voltage |
| AC or DC electric current measurement (use SELECT push button for switching to AC or DC). |
Part Buttons
| Push button | Measurement Function |
| SELECT | Switching AC or DC. Press the xanthous SELECT button to select alternate measurement functions on the rotary switch. |
| Display freezes present reading / press 2 sec to turn on LCD backlight. |
| REL | Relative zero mode |
RANGE | Manual or Automobile range switching. The default setting is Auto ranging, press to switch to manual ranging (selectable resolutions). Printing for two sec to render to car ranging. |
| MAX/MIN | Maximum / minimum reading memory. |
| Flash light |
Press
to enable the function when at relevant rotary switch office.
Auto Ability OFF
Car ability off: approx. fifteen minutes.
When the Meter is in automobile ability off way, printing any button to resume normal operation.
Measuring AC and DC Voltage
Printing SELECT button to select Air-conditioning/DC voltage measurement function.
To avoid personal injury or damage to the Meter, do non utilise
voltage higher than 600Vac and 600Vdc.
Measuring Air-conditioning and DC Current
Printing SELECT button to select AC or DC current measurement role.
To avoid personal injury or damage to the Meter:
- Do not attempt to make an in-excursion electric current measurement when the open-circuit potential to earth footing exceeding Air conditioning 600V or DC 600V
- Switch to proper function and range for your measurement.
- Do not identify the examination probe in parallel with a circuit when the examination leads are continued to the current terminals.
- Connect the test leads to the correct input A/mA μA current terminal and to the circuit before powering the circuit under test.
- For current range from viii-10A, do not measure current for more than 20 minutes. Wait for x minutes before taking another measurement
- After measurement, switching OFF the circuit'south power before removing test leads from the circuit.
Measuring Resistance
Disconnect circuit power and discharge all loftier-voltage capacitors before testing resistance.
Measuring Continuity
Disconnect circuit ability and discharge all high-voltage capacitors
before testing continuity.
Measuring Diode
Disconnect circuit power and discharge all high-voltage capacitors
earlier testing diode.
Measuring Capacitance
Disconnect circuit power and discharge all high-voltage capacitors before testing capacitance.
Measuring Frequency
To avoid personal injury or damage to the Meter, do not apply voltage higher than 600V.
Measuring Temperature °C / °F
-
To avoid personal injury or impairment to the Meter, do not apply thetemperature probe to any live conductive parts.
ii. Temperature sensor G type (nickel-chromium/nichrosi) thermocouple is suitable for temperature measurement below 230°C (446°F).
Measurement steps:
Step one: Plow the rotary switch to °C or °F position. The display will show "OL".
Step ii: Connect the temperature probe (K type) to the Meter and to the surface to be measured.
Non-Contact Voltage Sensing
- To avoid personal injury or damage to the Meter, do not test on uninsulated loftier voltage wires.
- Buzzer will sound and screen will display "OL" when sensing AC voltage between 90V and 600V
- Do not test on hazardous live wires higher than Air-conditioning 600V
- Before and afterwards hazardous voltage measurements, examination the Meter by approaching to a known source such as a line AC voltage or outlet to make up one's mind proper operation. See below figure.
Buzzer will sound when the detected voltage is ≥90V, and the buzzer will be on.The distances between the wire and the meter should be ≤8mm.
At NCV mode, LCD will brandish OL. No exam atomic number 82 connection are required for NCV measurement.
SPECIFICATION
Ambient temperature: 73.iv°F ±9 (23°C ±5°C); Relative temperature: ≤75%
Accuracy: ±(% of reading + digits)
Maximum voltage betwixt input concluding and globe ground: Ac 600Vrms or DC 600V
Fuse for mA µA input: F1 fuse, 0.5A H 1000V fast-fuse (6.3×32)mm
Fuse for 10A input: F2 fuse, 11A H 1000V fast-fuse, (ten×38)mm
Maximum display: Digital 3999 counts, updates 3/sec. Frequency: 9999 counts.
Analog pointer display: 41 segments, updates 30 times/sec. Over-range indication: OL
Range: Automatic and Manual
Altitude: Operating 2000m
Operating temperature: 0°C ~ +xl°C (32°F ~ 104°F)
Relative humidity: 0°C ~ +xxx°C (32°F ~ 86°F) ≤75%; +30°C ~ +twoscore°C (86°F ~ 104°F) ≤fifty%
Storage temperature: -10°C ~ +50°C (xiv°F ~ 122°F)
Electromagnetic compatibility: In an RF filed of 1V/m = Specified accuracy 5%
Battery: 9V, 6F22, NEDA1604 or equivalent Low battery indication:
Dimensions (L x W x H): 182 mm 10 ninety mm x 45 mm (7.2 in x three.5 in x 1.viii in)
Weight: Approx. 354g (0.78lb) with batteries installed
- DC Voltage Measurement
Range
Resolution
Accuracy
400.0mV
0.1mV ± (0.eight%+3LSD) 4.000V 1mV ± (0.viii%+1LSD)
40.00V
10mV
400.0V
100mV 600V 1V ±(ane.0%+3LSD)
Input impedance: About 10M ;
(Input impedance is £3Ge except DC 400mV range)
Overload protection: ±600V -
1. AC Voltage Measurement
Range
Resolution Accuracy 400.0mV
0.1mV ±(1.ii%+3LSD) 4.000V 1mV ±(1.0%+3LSD)
40.00V
10mV
400.0V
100mV
600V 1V ±(1.2%+3LSD)
Note: Manual range only for 400.0mV range.
Input impedance: Effectually 10MΩ
Frequency response: 45Hz ~ 400Hz
AM-520-EUR : Average sensing, rms indication.
AM-530-EUR: Truthful RMS.
Overload protection: 600Vrms - Resistance Measurement
Range
Resolution Accurateness 400.0W
0.1W ±(1.2%+2LSD) 4.000kW 1W ±(ane.0%+2LSD)
40.00kW
10W 400.0kW 100W
4.000MW
1kW ±(1.2%+2LSD)
40.00MW 10kW ±(1.five%+5LSD)
400Ω range: Measured value = (Measured display value – Short circuiting value of probe)
Open circuit voltage: Around 0.5V
Overload protection: 600Vrms -
: Circuit ON/OFFG : Diode measurement
Range
Resolution Accuracy
0.1Ω
Open circuit voltage is around 0.5V.
Resistance >150W, cablegram will not sound. Resistance £10W, buzzer will sound.
1mV
Display range is 0V to 2.0V. Normal voltage is around 0.5V to 0.8V for silicon PN junction.
Overload protection: 600V
- Capacitance Measurement
Range
Resolution Accuracy 40.00nF
10pF ±(three%+10LSD) under REL status 400.0nF 100pF ±(3%+5LSD) nether REL status
4.000μF
1nF xl.00μF 10nF ±(3%+5LSD)
400.0μF
100nF ±(4%+5LSD)
4000μF 1μF For reference only
Overload protection: 600V
- Frequency Measurement
Range
Resolution
Accurateness
10Hz~10MHz
0.01Hz~0. 01MHz ±(0.one%+4LSD)
Overload protection: 600Vrms
- DC Current Measurement
Range
Resolution
Accuracy
μA
400.0μA 0.1μA ±(1.0%+2LSD)
4000μA
1μA
mA
40.00mA
10μA
400.0mA
0.1mA
A
4.000A
1mA
±(1.2%+3LSD)
10.00A
10mA
Overload protection:
mA /µA range:F1 fuse, 0.5A H 1000V fast-fuse, ( 6.three×32)mm
x A range:F2 fuse, 11A H 1000V fast-fuse, ( 10×38)mm - Ac Electric current Measurement
Range
Resolution
Accuracy
μA
400.0μA 0.1μA ±(1.two%+3LSD)
4000μA 1μA
mA
xl.00mA
10μA
400.0mA
0.1mA
A
4.000A 1mA
±(1.five%+3LSD)
x.00A
10mA
Frequency response: 45Hz ~ 400Hz
AM-520-EUR : Average sensing, rms indication.
AM-530-EUR: True RMS.
Overload protection:
mA /µA range:F1 fuse, 0.5A H 1000V fast-fuse, ( half dozen.three×32)mm
10 A range:F2 fuse, 11A H 1000V fast-fuse, ( 10×38)mm - Temperature Measurement
Range
Resolution
Accuracy
-40 – 0°C
0.1°C @ < 400°C
i°C @ ³ 400°C
±(x%+iv°C)
>0 – 100°C
±(one.2%+3°C)
>100 – thou°C
±(2.5%+2°C)
-40 – 32°F
0.1°F @ < 752°F
1°F @ ³ 752°F
±(xx%+6°F)
>32 – 212°F
±(1.8%+half-dozen°F) >212 – 1832°F ±(2.5%+iv°F)
Overload protection: 600V
K blazon (nickel-chromium/nichrosi) thermocouple must be used for temperature measurement.
MAINTENANCE AND REPAIR
If the Meter fails to operate, check battery, exam leads, etc., and supersede as necessary.
Double check the followings:
- Replace the fuse or battery if the meter does not work.
- Review the operating instructions for possible mistakes in operating procedure.
Quick cheque on 0.5A FUSE:
Step one: Turn the rotary switch to Ω function.
Step ii: short-circuit E/V/Ω/Hz terminal and mA/μA terminal.
Resistance reading ≤ 1MΩ: the fuse is OK
Resistance reading "OL": the fuse is open. Replace the fuse every bit specified.
Quick check on 10A FUSE:
Footstep 1: Plow the rotary switch to Ω function.
Footstep ii: short-circuit E/V/Ω/Hz terminal and A last.
Resistance reading ≤ 0.5Ω: the fuse is OK.
Resistance reading "OL": the fuse is open. Replace the fuse as specified.
Except for the replacement of the bombardment, repair of the meter should be performed only by a Factory Authorized Service Center or by other qualified instrument service personnel.
The front end console and case can be cleaned with a mild solution of detergent and h2o. Apply sparingly with a soft material and allow to dry completely before using. Do not utilise aromatic hydrocarbons, Gasoline or chlorinated solvents for cleaning.
BATTERY AND FUSE REPLACEMENT
Warning
To avoid shock, injury, or harm to the Meter:
Disconnect test leads earlier opening case.
Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings specified.
Replacing BATTERY follow below steps:
- Disconnect the test lead probe from measuring circuit.
- Plow the Meter to OFF position.
- Remove the screws from the bombardment encompass and open the battery cover
- Remove the batteries and replace with one 9V (6F22) or equivalent. The battery cover provides the correct polarity plumbing fixtures construction design.
Install the battery in the battery cover. - Put the bombardment embrace back and re-fasten the screw. Bombardment: 9V (6F22) Bombardment or equivalent
Replacing FUSE follow below steps:
- Disconnect the exam atomic number 82 probe from measuring circuit.
- Plough the Meter to OFF position.
- Remove the screws from the enclosure and open the enclosure.
- Remove the broken fuse and supercede with new specified fuse.
- Put the enclosure back and re-fasten the screw.
Fuse ratings:
mA /µA input terminal: F1 fuse, 0.5A H 1000V fast-fuse, ( six.iii×32)mm
ten A input final: F2 fuse, 11A H 1000V fast-fuse, ( 10×38)mm